Journal of Shanghai Jiao Tong University (Medical Science) >
Comparison of the central positioning ability of three single-file nickel-titanium instruments in simulated curved root canals
Received date: 2022-01-05
Accepted date: 2022-04-10
Online published: 2022-05-28
Supported by
Research Discipline Fund from Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine, and College of Stomatology, Shanghai Jiao Tong University(KQYJXK2020)
·To compare the ability of VDW Reciproc Blue (Blue), FKG XP-endo shaper (XP) and M3-L platinum (M3) instruments to maintain the original curvature of root canal (i.e. central positioning ability) in preparing curved root canals.
·Thirty resin-simulated curved single root canals were randomly divided into three groups. Root canal preparation was performed with a Blue file, an XP file or an M3 file randomly. Red or blue-black ink was injected into the root canal before and after preparation and photographs were taken under the stomatology surgical microscope. The images before and after preparation were overlapped. The apex of the prepared root was set as a circle center and the radius started at 1 mm. A concentric circle was made every 1 mm increased and sequentially incremented to 10 mm so that it intersected the image to form 10 measurement sites, and then the removal of ten measurement sites on the inside and outside of the root canal was recorded for statistical analysis. After the root canal was prepared the file was observed for tensile deformation under the stomatology surgical microscope.
·There was significant difference in the absolute value of the difference between the internal and external removal of root canal between the M3 group and the Blue group, as well as the M3 group and the XP group (P=0.003 1, P=0.002 0); but there was no significant difference between the Blue group and the XP group (P=0.350). The file body of XP group was observed to have de-helix deformation under the stomatology surgical microscope, which did not return to its original shape when placed in hot water, and no deformation was observed in the file body of M3 group and Blue group.
·The three kinds of single-file nickel-titanium instruments have good formation ability when preparing the curved root canals. But XP file and Blue file can maintain the original curvature of root canals better than M3 file. XP files are prone to deformation when preparing resin root canal, therefore the use of XP should be appropriately reduced clinically.
Tingting LI , Yueping PAN . Comparison of the central positioning ability of three single-file nickel-titanium instruments in simulated curved root canals[J]. Journal of Shanghai Jiao Tong University (Medical Science), 2022 , 42(5) : 690 -694 . DOI: 10.3969/j.issn.1674-8115.2022.05.020
1 | SCHNEIDER S W. A comparison of canal preparations in straight and curved root canals[J]. Oral Surg Oral Med Oral Pathol, 1971, 32(2): 271-275. |
2 | SCH?FER E, DIEZ C, HOPPE W, et al. Roentgenographic investigation of frequency and degree of canal curvatures in human permanent teeth[J]. J Endod, 2002, 28(3): 211-216. |
3 | ANDREASEN G F, BRADY P R. A use hypothesis for 55 Nitinol wire for orthodontics[J]. Angle Orthod,1972, 42(2): 172-177. |
4 | WALIA H M, BRANTLEY W A, GERSTEIN H. An initial investigation of the bending and torsional properties of nitinol root canal files[J]. J Endod, 1988, 14(7): 346-351. |
5 | HAAPASALO M, SHEN Y. Evolution of nickel-titanium instruments: from past to future[J]. Endod Top, 2013, 29(1): 3-17. |
6 | RUDDLE C J, MACHTOU P, WEST J D. The shaping movement: fifth-generation technology[J]. Dent Today, 2013, 32(4): 94, 96-94, 99. |
7 | KESKIN C, SAR?Y?LMAZ E, DEMIRAL M. Shaping ability of Reciproc Blue reciprocating instruments with or without glide path in simulated S-shaped root canals[J]. J Dent Res Dent Clin Dent Prospects, 2018, 12(1): 63-67. |
8 | SINGBAL K, JAIN D, RAJA K, et al. Comparative evaluation of apically extruded debris during root canal instrumentation using two Ni-Ti single file rotary systems: an in vitro study[J]. J Conserv Dent, 2017, 20(2): 64-67. |
9 | 辛悦, 杨健. XP-endo系列镍钛锉的研究进展[J]. 口腔医学研究, 2019, 35(6): 521-523. |
9 | XIN Y, YANG J. Advancement in XP-endo series nickel-titanium instruments [J].J Oral Sci Res, 2019, 35(6): 521-523. |
10 | ALOMARI T, MUSTAFA R, AL-FODEH R, et al. Debris extrusion using reciproc blue and XP endo shaper systems in root canal retreatment[J]. Int J Dent, 2021, 2021: 6697587. |
11 | 冯海玲, 史璐. 四种往复运动单支锉系统的研究进展[J]. 口腔医学研究, 2020, 36(2): 111-113. |
11 | FENG H L, SHI L. A review of four reciprocating single-file systems[J]. J Oral Sci Res, 2020, 36(2): 111-113. |
12 | 张琦, 张旗. XP-endo Shaper和ProTaper Next对弯曲根管碎屑清理效果和根尖碎屑推出量的比较[J]. 临床口腔医学杂志, 2021, 37(8): 469-471. |
12 | ZHANG Q, ZHANG Q. Comparative study of XP-endo Shaper and ProTaper Next on intracanal debris removal and apical debris extrusion in curved root canals[J]. J Clin Stomatol, 2021, 37(8): 469-471. |
/
〈 |
|
〉 |