透视立体影像分析技术在假体无菌性松动检测中的应用研究
杨涵, 雷浩, 徐彼得, 吴淏, 马寻君, 皇艳波, 毛远青, 张经纬, 王金武

Application of fluoroscopic stereophotogrammetric analysis in the detection of aseptic loosening of prostheses
YANG Han, LEI Hao, XU Bide, WU Hao, MA Xunjun, HUANG Yanbo, MAO Yuanqing, ZHANG Jingwei, WANG Jinwu
表1 2种工况下算法1和算法2误差对比
Tab 1 Error comparison of algorithm 1 and algorithm 2 under different working conditions
ItemAlgorithm 1Algorithm 2t valueP value
Working condition 1
ry /(°)6.06±5.953.79±3.362.3520.021
ty /mm1.91±1.150.20±0.1710.4360.000
Working condition 2
ry /(°)4.15±3.183.47±2.791.1350.259
ty /mm2.68±1.950.34±0.288.3960.000