›› 2010, Vol. 30 ›› Issue (10): 1307-.doi: 10.3969/j.issn.1674-8115.2010.10.030

• Technique and method • Previous Articles     Next Articles

Assessment of effect and development of percutaneous pulmonary artery thrombus clear catheter

FAN Ji-hai1, GU Xiu-lian1, CHAO Sheng-wu1, ZHANG Peng1, CHEN Yue-guang2, MI Zhi-cai3   

  1. 1.Department of Cardiology, the 455th Hospital of People's Liberation Army, Shanghai 200052, China;2.Department of Cardiology, Minhang Branch of Ruijin Hospital, Shanghai Jiaotong University School of Medicine, Shanghai 201100, China;3.Kangge Medical Equipments Limited Company, Shanghai 200052, China
  • Online:2010-10-25 Published:2010-10-27
  • Supported by:

    Foundation of Changning District Health Bureau of Shanghai, 20074Y16001

Abstract:

Objective To develop a percutaneous pulmonary artery thrombus clear catheter used in patients with acute pulmonary thromboembolism (APTE) and to assess its effect. Methods The commonly used clinical catheter materials were selected under ultrasound welding, film coating, and the hot working technologies to make a selfexpandable nitinol umbrella mounted at a catheter-tip to suck thrombus. In vitro and in vivo tests were completed on the established sheep APTE models (n=10) for testing the safety, efficiency, and operability of the device. Results The device were made successfully along with iterative improvements. The tests revealed that thrombus could be extracted perfectly followed by partial recanalizations of pulmonary artery and significant improvements of hemodynamic parameters, such as pulmonary artery pressure, peripheral artery pressure, arterial partial pressure of oxygen, and heart rate. Conclusion The device is safe and feasible in sheep APTE models, and it is confident to have a clinical trial with further improvements of the device.

Key words: acute pulmonary thromboembolism, thrombectomy, catheterization, interventional therapy